The short version. To verify a unit, once a year or any time: Daily Counter Checks, 2 Daily QC, then a background count. Run Factory Recalibration only when Daily QC recommends it, or when LTI asks you to. It is a very aggressive recalibration of every part of the counting circuitry, it erases all stored efficiencies, and it can take hours with a weak source.

What Factory Recalibration does

It rebuilds the energy calibration of every well from scratch: high voltage, zero, lower level and each well's gain. Because that changes everything to do with the detectors, it also returns the unit to factory settings for anything to do with counting. It erases every stored isotope efficiency, every background value and the spilldown (crosstalk) factors. All of them must be re-measured before the unit is used for wipes again. That is why the counter shows three warning screens before it starts, and why the last one needs a different key to proceed.

What to run instead

  • Daily QC (Daily Counter Checks, 2 Daily QC) counts a known check source and compares each well's efficiency with the stored value. A well within about 5% passes. This is the check that confirms the calibration is good.
  • Background (Daily Counter Checks, 1 Background Check) measures what each well counts with nothing in it. Count it at least every 24 hours and after any power-off, for 10 minutes or longer.
  • Chi Square and FWHM (System Diagnostics, 1 and 2) check counting statistics and detector resolution. Both need a matched set, one source in every well.

When Factory Recalibration is needed

Only when the counter fails its Daily QC checks and the instrument recommends recalibrating, or when LTI service asks you to. On a healthy unit that is typically only after several years in the field.

Co-57 or Cs-137

  • Co-57 (122 keV) is a good, solid calibration, and works best for departments primarily using Tc-99m (140 keV), because it lines the scale up right where they count most. It is also far faster. Linearity on these counters is good, so after a Co-57 calibration the unit can count any isotope.
  • Cs-137 is best for departments using multiple isotopes. As well as the 662 keV peak, it uses the barium x-ray peak at 32 keV to set the low end of the scale, which gives superior linearity across the whole isotopic range. That extra step is why it takes longer and needs a stronger source.

One source, or one in every well

How the calibration runs depends on the source's Reference setting in the Isotope Library (6 Isotope Library, 71 EDIT ISO). Set to Multi, it expects a matched set, one source in every well at the same time. Set to Single, it walks one source through the wells and tells you when to move it. A Cs-137 rod always runs as Single. If the unit expects a matched set and only one well has a source, the other wells see nothing and the calibration keeps hunting.

Source strength

For Cs-137, LTI recommends at least 0.5 µCi. On a healthy detector, 0.1 µCi (100 nCi) is the lowest that works, and only while the source is fairly new: Cs-137 loses about 2% a year, so a rod that started at 0.1 µCi falls below that within a few years. Co-57 calibrators typically ship at about 150,000 DPM. See Choosing a calibration source.

How long it takes

The time depends almost entirely on the source. Each well has to collect a set number of counts at several stages (140,000 per pass with Cs-137, 65,000 with Co-57), and every adjustment restarts the count. Approximate times on a 4-well unit:

SourceWhole calibration
Co-57, matched setabout 10 minutes
Co-57, single sourceabout 15 to 30 minutes
Cs-137, 0.5 µCiabout 1 hour
Cs-137, 0.1 µCi, new2 to 3 hours or more
Cs-137, 0.1 µCi, several years oldlonger still, and it may not complete

Let it run. If one stage on one well runs far past these times with no change, or a WARNING says a well has been disabled, it will not finish on its own. Stop and contact LTI, rather than repeating the calibration.

Step by step (Multi-Wiper™ 3.5.1)

  1. Turn back on any well that was disabled by hand: 5 System Setup, 5 Detector Setup.
  2. Check the source's Reference setting (Multi or Single) in the Isotope Library.
  3. 7 System Diagnostics, 10 Factory Recal, then 1 Energy Cal. (2 Energy Cal, FWHM, ChiSq needs a matched set.)
  4. Read all three warning screens. The third one needs 7 to proceed.
  5. Choose Co-57 or Cs-137, then follow the prompts. With a single source, move it to each well when the screen says "Move ref source to: Well n" and press #.
  6. When it finishes, take the source out, store it well away from the counter, and count a background.
  7. Re-measure every stored efficiency: 6 Isotope Library, 71 EDIT ISO, the isotope's number, then Measure Efficiency. Repeat for every isotope you use.
  8. Print the summary (Calibration, 3 Print Last Cal). It shows the new settings next to the previous ones.