What well-to-well spread is
On a multi-well counter, the system efficiency check counts the same calibrated source in every well and reports each well’s efficiency, then the spread — the gap between the highest and lowest well. A small spread is expected and healthy. The question is never “is there any spread” — there always is — but “is this spread telling me something.”
Cause one: the statistical floor
Every count carries an irreducible statistical uncertainty set by the number of events you collect. Collect 120,000 counts and the one-sigma noise on that measurement is about 0.3%. Collect only 10,000 counts and it climbs to about 1%. Collect a couple of thousand and you are at 2–3% before the instrument has done anything at all.
This matters enormously for low-yield isotopes. A tracer like Cr-51 emits a counted gamma in under 10% of its decays, so at clinical activities the effective count rate is low — and a 4–5% well-to-well spread can be entirely statistical noise, with a perfectly healthy instrument underneath. The fix is not service; it is more counts: a longer count time, or a higher-yield source such as Cs-137 for the stability check.
Cause two: normal photomultiplier jitter
A healthy photomultiplier does not hold its gain to the last decimal. Its photopeak drifts a little from day to day, and a detector settles over its first year of service as the tube finds its long-term operating point. That movement is normal and within specification. It only becomes visible as “efficiency drift” when something amplifies it — most often an energy window that has been narrowed below the factory default, which turns a small, healthy peak shift into a larger swing in counts. (See Energy windows — why a properly set window beats a tight one.)
How to tell real drift from normal scatter
Four checks separate a fault from the noise floor:
- Count more. Repeat the check with far more total counts — a longer count time, or a higher-yield isotope. If the spread shrinks toward a fraction of a percent, it was statistics. If a high-statistics check still shows several percent, now you have something real to look at.
- Look at the trend, not the snapshot. Plot each well over weeks. Random scatter inside a stable band is noise. A monotonic walk in one direction, or a clear step that then persists, is the signature of a genuine change.
- Check the photopeak position. If a well’s efficiency moves together with its photopeak channel, you are watching gain move. Whether that matters depends on size and persistence — a few keV that wanders is normal; a steady march is not.
- Rule out the window. Confirm the energy windows are at the factory default before trusting the spread at all. A narrowed window manufactures apparent drift from normal motion.
When a spread is actually worth acting on
Act when the spread is real (it survives a high-statistics count), localized (one well, not the whole array), and progressive (it grows or steps over time rather than scattering around a stable mean). A single well that walks steadily away from the group, or throws repeatable excursions a high-count check still shows, has earned a closer look. A spread that is none of those things — that shrinks when you count longer and otherwise drifts randomly within a band — is the instrument behaving exactly as it should.
What to do
- Use a high-yield source for stability checks. Cs-137 gives you the counts to push the noise floor down to a fraction of a percent, so a real change stands out.
- Confirm your windows are at factory default before reading anything into a spread.
- Trend each well over time. The shape of the history tells you far more than any single reading.
- Don’t benchmark against your quietest tube. Judge each well against specification, not against the best-behaved detector in the array.